ASTM F1249 Water Vapor Transmission Rate

ASTM F1249 Water Vapor Transmission Rate

ASTM F1249: Water Vapor Transmission Rate of Flexible Materials

Gravimetric techniques involve weighing a sample and waiting. For a metallic film or coated barrier film, the weight change after a week may fall within the balance’s noise level, so you never get a reliable answer. ASTM F1249 uses infrared detection to measure water vapor in a carrier gas, achieving transmission rates many times lower than those measurable by weighing, and in a much shorter time frame.
MatX Lab performs WVTR testing on packaging films, laminates, coated substrates, and barrier structures.

How the Measurement Works

The specimen is clamped between two halves of a diffusion cell, sealing them together.
One side holds a controlled humidity, generated by a wetted sponge or a humidified gas stream. The other side is swept with dry nitrogen. Water vapor permeating through the film joins the nitrogen and is carried to a modulated infrared sensor, which responds to the water vapor concentration in the stream.
There is continuous product output. This means there is no need to measure the mass difference after two weighings; an increasing signal marks the process due to permeation until it reaches a stable value. Thus, the equilibrium is reached visibly rather than hypothetically.

Standard Conditions

 

Parameter Typical value
Test temperature 37.8 °C (100 °F), other temperatures on request
Driving humidity 90% RH or 100% RH depending on the configuration
Carrier gas Dry nitrogen at controlled flow
Reported unit g/(m²·day) at the stated conditions
Test area Commonly 50 cm², masked down for high barrier films
Detector Modulated infrared sensor

 

The 37.8 °C and high humidity combination is the packaging industry convention, chosen to accelerate permeation while staying below temperatures that would alter the film. Results at these conditions do not transfer directly to ambient or chilled service, since permeation is strongly temperature-dependent.

Sensitivity and Its Limits

This test has broad range coverage. Ordinary polyolefin films pass a few grams per square meter per day, and the tests can be completed in no time. Barrier films and metalized films pass very low amounts, but high-barrier films require many days to test.
Two factors influence the lowest level measurable in practice. The first is system background—the amount of vapor reaching the sensor through paths other than the sample film. The second is test duration, since very slow permeation requires more time to distinguish a true plateau from instrument drift.
A calibration film with a known transmission rate verifies instrument response, and running one alongside a barrier program separates instrument issues from material results.

What Goes Wrong

Choosing Between F1249 and Gravimetric Methods

 

ASTM F1249 Gravimetric, ASTM E96
Detection Infrared sensor on carrier gas Mass change of a sealed dish
Best suited to Flexible films, laminates, barrier structures Building materials and moderate barriers
Sensitivity Reaches low transmission rates Limited by balance resolution and internal air gap
Duration Hours for ordinary films Days to weeks
Sample form Flat flexible specimen clamped in a cell Sheet or board sealed to a dish

 

Oxygen transmission on the same films is measured separately under ASTM D3985, and most barrier programs need both, since a structure can perform well against one gas and poorly against the other.

Specimen Requirements

Flatten film or laminate that is oversized for the cell and interleaved for shipping. A crease intersecting the test area is a defect.
Determine the product contact surface. Coated and laminated films are non-symmetric and may show varying transmission depending on the measurement direction.
Send enough for replicates plus spares. On high barrier material, one anomalous specimen is usually a defect rather than the material, and distinguishing the two needs more than one result.

 

ASTM F1249 FAQs

Why is weighing a dish not enough?

The first reason is that mass changes are very minute in barrier film transmission. Second, air in the dish adds resistance in series with the sample.

How long does a barrier film take?

Ordinary films reach steady state within hours. High barrier structures can take a day or more, and reporting before the signal plateaus produces a value that understates transmission.